À propos de : Dipstick measurements of urine specific gravity are unreliable        

AttributsValeurs
type
Is Part Of
Subject
Title
  • Dipstick measurements of urine specific gravity are unreliable
has manifestation of work
related by
Abstract
  • AIM. To evaluate the reliability of dipstick measurements of urine specific gravity (U-SG). METHODS. Fresh urine specimens were tested for urine pH and osmolality (U-pH, U-Osm) by a pH meter and an osmometer, and for U-SG by three different methods (refractometry, automatic readout of a dipstick (Clinitek-50), and (visual) change of colour of the dipstick). RESULTS. The correlations between the visual U-SG dipstick measurements and U-SG determined by a refractometer and the comparison of Clinitek®-50 dipstick U-SG measurements with U-Osm were less than optimal, showing very wide scatter of values. Only the U-SG refractometer values and U-Osm had a good linear correlation. The tested dipstick was unreliable for the bedside determination of U-SG, even after correction for U-pH, as recommended by the manufacturer. CONCLUSIONS. Among the bedside determinations, only refractometry gives reliable U-SG results. Dipstick U-SG measurements should be abandoned.