Attributs | Valeurs |
---|
type
| |
Is Part Of
| |
Subject
| |
Title
| - Evaluation of repeatability for the automatic estimation of endothelial cell density in donor corneas
|
has manifestation of work
| |
related by
| |
Abstract
| - Background/Aims:. The repeatability of an automatic system for estimation of endothelial cell density (ECD) from microscopy images in donor corneas was assessed. Methods:. A computer program for the automatic ECD estimation from frequency analysis was previously presented. An evaluation of its repeatability was performed on a data set containing 200 corneas by assessing the accuracy and precision of automatic versus manual values. For each cornea, 2-21 images (1536 total) at 100× for automatic ECD and one image at 200× for manual ECD were available. Results:. Accuracy of automatic ECDs was −45 (SD 99) cells/mm2 (−2% (4%)). Precision of repeated automatic ECDs on the same cornea was 62 (30) cells/mm2 (3% (1%)). The algorithm was also capable of identifying all non-processable images. Conclusion:. The proposed automatic technique proved to be reliable for its routine use in a typical eye bank setting like the one considered here.
|
article type
| |
publisher identifier
| |
Alternative Title
| - Repeatability of automatic ECD estimation
|
is part of this journal
| |
PubMed ID
| |