Documentation scienceplus.abes.fr version Bêta

À propos de : Evaluation of the BacT/ALERT PZA kit in comparison with the BACTEC 460TB PZA for testing Mycobacterium tuberculosis susceptibility to pyrazinamide        

AttributsValeurs
type
Is Part Of
Subject
Title
  • Evaluation of the BacT/ALERT PZA kit in comparison with the BACTEC 460TB PZA for testing Mycobacterium tuberculosis susceptibility to pyrazinamide
has manifestation of work
related by
Author
Abstract
  • Objectives. To compare the performance of the BacT/ALERT PZA kit (BioMerieux, Marcy l'Etoile, France) with the radiometric BACTEC 460TB PZA test (Becton-Dickinson method) for testing Mycobacterium tuberculosis susceptibility to pyrazinamide. Methods. A total of 50 M. tuberculosis strains were tested. Thirty of these strains had been previously considered pyrazinamide-susceptible and 20 pyrazinamide-resistant by BACTEC 460TB. Results. Final overall agreement was 100%. Time needed for the susceptibility test was 6.69 days for the BacT/ALERT PZA kit versus 8.07 days for the BACTEC 460TB PZA test. Conclusions. BacT/ALERT PZA test is an excellent alternative to BACTEC 460TB for pyrazinamide susceptibility testing.
article type
publisher identifier
  • dkm252
Alternative Title
  • Aragón et al.
  • BacT/ALERT PZA in comparison with BACTEC 460TB PZA
is part of this journal



Alternative Linked Data Documents: ODE     Content Formats:       RDF       ODATA       Microdata